Intended use
This product is not intended to be used by the patient alone. Lifting and transferring a patient shall always be performed
with the assistance of at least one caregiver. This product is used as a means to perform the lift but is not in contact with the
patient; therefore we do not go into the various patient conditions in this manual. Contact your Hill-Rom representative for
support and advice.
Certain environments and conditions can limit the correct us of the mobile lifts, including:
Thresholds, unlevel floor surfaces, various obstacles, and extra-thick carpets. These environments and conditions can cause the
wheels of the mobile lift not to roll as intended, possible imbalance in the mobile lift, and increased exertion by the caregiver. If
you are uncertain that your care environment fulfils the requirements for correct use of the mobile lift, please contact your Hill-
Rom representative for further advice and assistance.
Before use, make sure that:
• the lift is assembled in accordance with the assembly instructions;
• the lifting accessory is properly attached to the lift;
• the battery has been charged for at least 6 hours;
• you have read the instruction guides for the lift and lifting accessories;
• personnel using the lift are informed of the correct operation and use of the lift.
Before lifting, always make sure that:
• the lifting accessories are not damaged;
• the lifting accessory is correctly attached to the lift;
• the lifting accessory hangs vertically and can move freely;
• the lifting accessory is selected appropriately, in terms of type, size, material and design, with regard to the patient’s needs;
• the lifting accessory is correctly and safely applied to the patient in order to prevent injuries;
• the latches are intact; missing or damaged latches must always be replaced;
• the sling’s strap loops are correctly connected to the sling bar hooks when the sling straps are stretched up but before the
patient is lifted from the underlying surface.
Viking™ M mobile lift have been tested by an accredited testing institute and comply with the requirements specified in the
Medical Device Directive 93/42/EEC, for Class I products.
Viking™ M mobile lift comply with the requirements in EN ISO 10535:2006 and CAN/CSA Z10535-03 and applicable parts of
IEC 60601-1, EN 60601-1-2, ANSI/AAMI ES60601-1, CAN/CSA C22.2 no 60601-1, and AS/NZS 3200-1-0.
No modification of this product is allowed.
Particular care must be taken when using powerful sources of electromagnetic disturbance, such as diathermy, etc, so that,
for example, diathermy cables are not positioned on or near the lift. If you have any questions, please consult the responsible
assistive device technician or the supplier.
The lift may not be used in areas where flammable mixtures may occur, for example, in areas where flammable goods are
stored.
Unbalanced lifting poses a tipping risk and may damage the lift equipment!
Do not raise the lift arm manually!
Never leave a patient unattended during a lifting situation!
Safety Instructions
This Caution notice is found on the Battery:
CAUTION!
NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL
DO NOT SHORT CIRCUIT
USE THE SPECIFIED CHARGER ONLY
MAY EXPOLODE IF DISPOSED IN FIRE
This Caution notice is found on the Control box:
CAUTION!
NOT TO BE OPENED BY UNAUTHORIZED PERSONNEL