www.liko.com
2.0
Care and Maintenance
Checklist
For safe and troublefree operation, there are a few routine measures which should be performed
every day the lift is used.
• Inspect the lift and check for any external damage.
• Check the connecting joints of the sling bar.
• Check the lift strap for wear and that the strap is not twisted.
• Check the function of the safety latch.
• Check the control function of the lift movement.
• Check that the lifting height is correctly adjusted and that the emergency-lowering device works.
• Charge the batteries every day the lift is used, and check that the battery charger works.
As needed, clean the lift with warm water or disinfectant. Caution! Do not use cleaning products
that contain phenol or chlorine, since it can damage the aluminum and polyamide material.
Inspection and Control
Likorall should be thoroughly inspected for wear of parts at least once per year. Inspection should
be in accordance with Liko service instructions and conducted by authorized individuals.
WARNING!
Use only authorized Liko service personnel and Liko spare parts.
Transport and Storing
During transportation, or when the patient lift is not to be used for some time, the emergency
button should be pushed in. When storing, keep the patient lift at a temperature exceeding free-
zing point and at a relative humidity no more than 60%.
We always try to improve our products. Naturally, this means changes from time to time.
Please contact our local representative or visit www.liko.com to receive the latest updates.