Service Manual Golvo™ 8000/8008
Mobile Lift
• 3EN400404 Rev. 6 20
www.liko.com
l
www.hill-rom.com
6.2 Service, Control Unit CBL - Reset Service Interval Parameters
INPACT ON PARAMETERS
Changes of parameters when LSS, LDS or Manuel Service
reset has been used:
Parameter:
1. LSS:
2. LDS:
3. Manual Service:
Check Points in PM ***
Service Light *
Unlit
Unlit
Unlit
TAAxT
unchanged
unchanged
unchanged
x ***
TAAxTR
unchanged
unchanged
unchanged
PAAxT **
0
0
0
x ***
PAKT **
0
0
0
TADT
unchanged
unchanged
unchanged
AAS
unchanged
unchanged
unchanged
x ***
ASERV
+1
+1
unchanged
x ***
IMED
unchanged
unchanged
unchanged
x ***
ALA
unchanged
unchanged
unchanged
x ***
AALA
unchanged
unchanged
unchanged
x ***
AOLA
unchanged
unchanged
unchanged
x ***
AABT
unchanged
unchanged
unchanged
x ***
AEUP
unchanged
unchanged
unchanged
x ***
AFSERV
0
0
+1
x ***
AEDWN
unchanged
unchanged
unchanged
x ***
SFSTAT
N/A
N/A
N/A
IDNR
unchanged
unchanged
unchanged
x ***
AOLUP
unchanged
unchanged
unchanged
x ***
PADT **
0
0
0
x ***
IMED20
unchanged
unchanged
unchanged
ELT
unchanged
unchanged
unchanged
ELTCNT
unchanged
unchanged
unchanged
CLIM1
unchanged
unchanged
unchanged
CLCT
unchanged
unchanged
unchanged
CLIM2
unchanged
unchanged
unchanged
SINTT
unchanged
unchanged
unchanged
SINTAS
unchanged
unchanged
unchanged
CYCLET
unchanged
unchanged
unchanged
PAOLUP **
0
0
0
x ***
* Indicates Service Periodic Inspection interval reached.
** Service Interval Parameters (the PAAxT, PAKT, PADT and PAOLUP are all used to determine Service intervals)
*** Parameters to use in the Preventative Maintenance
CBL RESET SERVICE INTERVAL PARAMETERS
After the Periodic Inspection has been performed Service Light and Service Interval Parameters*
in the
Liko Diagnostic
System™ has to be reset. Reset can be done using
one
of the following
Service Tools/Methods:
Prod. No.
21090024
1. Liko™ Service System, LSS
(
Requires a computer)
Read and store the data from the Control Unit CBL into a computer. Data can also be
transfered to a Liko™ Server. Can also reset the ELT-light (Expected Life Time).
Use when Periodic Inspection or Preventative maintenance is performed.
Service Documents for the lift can be printed. Use for tracking and and transfer data
when Spare Parts are changed.
Prod. No.
21090016
2. LDS Service (LDS)
Reads the data from the Control Unit CBL. Use when Periodic
Inspection or Preventative
maintenance is performed.
-
3. Manual Service
Resets the Service Interval Parameters*.
Service Interval Parameters should only be reset after a complete Periodic Inspection
Use the following combination of buttons on the control panel:
A: Press and hold button 1 & 2
B: While holding 1 & 2, press button 3, and hold all three
buttons for three seconds.
Works from Item No: CBL20001-05, see
label on CBL
1
2
3