Service Manual Golvo™ 8000/8008
Mobile Lift
• 3EN400404 Rev. 6 22
www.liko.com
l
www.hill-rom.com
8.0 Periodic Inspection Information,
Golvo™ 8000/8008 mobile lift
According to ISO 10535:2006 Annex B, Periodic Inspection of medical device Liko™ Mobile Lift should be
performed at least once a year. For Liko™ mobile lifts Periodic Inspection Liko Mobile lifts (3EN371001) is
available in Agile and on the Partner Web.
Liko™ lifts must be thoroughly inspected when the Service Light at the Control Unit illuminates (if applicable)
or at least once per year. Inspection and service must be carried out by Hill-Rom authorized personnel.
Periodic Inspection
Liko™ mobile lifts
Liko™ sit-to-stand lifts
3EN371001 Rev. 1
Valid for following lifts: Viking™, Golvo™, Uno™, Liko™ M220/230, LikoLight™, Sabina™, RollOn™.
Control Unit CBL
If Liko Service System (LSS) or Liko Diagnostic System (LDS) are available, data can be read
from lifts equipped with Control Unit CBL. LSS store data and LDS show data, write down
the values in the boxes below.
TAAxT
Ah
Control Unit Jumbo Care
From the display, write down the
values in the boxes below.
As
BATTERY
Total work, lifting actuator
To track how the lift has been used, write down the values from the control units (if applicable):
A S
Expected life time:
Viking™ M mobile lift
700 Ah
Viking L
mobile lift
910 Ah
Viking XL mobile lift
1020 Ah
Golvo 8/88 mobile lift
700 Ah
Expected life time:
Viking™ M mobile lift 2 520 000 As
Lift information
Customer reference
Lift:
Agreement No:
Product Number:
Name:
Serial Number:
Address:
Production year:
Post Code
Inspection sign off
Complete inspection according to Instructions for the check points (page 3-9), fill in page 2 and sign below.
Approval to use the mobile lift
Approved
Not approved
If the mobile lift has one or more inspection points “Not approved”, the lift must not be used.
To be actioned:
Actions according to the “INSPECTION ITEMS NOT APPROVED” should be performed immediately. After performed actions
sign below. If anything is unclear or if you have questions, please contact Hill-Rom or your local Hill-Rom representative.
Contact information is to be found at www.hill-rom.com.
Inspection performed
by:
Date:
Next inspection:
Inspection performed in accordance with ISO 10535:2006 Annex B- Periodic inspection
In this document, this warning symbol indicates that special care should be taken.
If instructions are not followed there is a risk of serious injury.
Utgått från PI:
Viking M, L XL
kollat igenom punkterna i övriga:
Golvo 8/88
Golvo 7/77
Uno 200
Uno 100-102
Uno 102 EE
Liko M220/230
LL
Viking S/XS
Sabina
RollON
1(9)
Liko technical documentation can be found under “My information” - “Documents” on the Partnerweb:
https://partner.liko.com/ If you don’t have access, contact Ann-Christin Sundström for registration:
E-mail: [email protected], Tel: +46 (0)920 - 47 47 10
Periodic Inspection may only be performed by personnel authorized by Hill-Rom and using
original Liko™ spare parts.
Available Liko Periodic Inspections:
•
3EN371001 Periodic Inspection Liko Mobile lifts
•
3EN111001 Periodic Inspection Liko OH Rail system
•
3EN191001 Periodic Inspection Overhead Lifts
• 3EN301001 PI Freespan Freestand
•
3EN601001 Periodic Inspection Liko accessories
Also available is Periodic Inspection Slings. This inspection should be performed by a person who is
suitable, has good experience and sufficient knowledge about design, use and maintenance of the sling.
•
7EN160834 General Periodic Inspection Protocol Liko Slings and Sabina SupportVests
NOTE!
Periodic Inspection shall be performed at least once per year.
The accumulated use of the lift (As) should be logged at each
Periodic Inspection!