background image

9

www.hillrom.com

7EN140108 Rev. 4  •  2020

Recommended separation distances between portable and mobile RF 

communications equipment and the mobile lift

The mobile lift is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The 

customer or the user of the mobile lift can help prevent electromagnetic interference by maintaining a minimum distance 

between portable and mobile RF communications equipment (transmitters) and the mobile lift as recommended below, 

according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter

(W)

Separation distance according to frequency of transmitter

  (m)

150 kHz to 80 MHz

P

d

2

,1

=

80 MHz to 800 MHz

P

d

2

,1

=

800 MHz to 2.7 GHz

P

d

3

,

2

=

0.01

0,12

0,12

0,24

0.1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) 

can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power 

rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1:

 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec

-

tion from structures, objects and people.

•  L

ift mast with control box and hand control, sling bar with 

latches, 2 M6 screws

• 

Armrest

•  B

ase, with width adjustment motor

 Lock the wheels on the base before starting to assemble the lift.

2.  Screw the two provided M6 screws into the upper 

holes on the lift mast.

  No screws in the lower holes!

1.   A)  Position the lift mast between the two black plastic plugs on 

the base cross member.

  B)  Then push the mast forwards as in the figure above so that 

it hooks onto the cross member.

Assembly

Before assembly, make sure you have the following parts and tools:

•  Battery

•  T

ools: 4, 5 mm hex keys

• 

 

Bag containing: charger cable, charger connector cable

•  

Instructions for use

A

B

Summary of Contents for 2000045

Page 1: ...mance and safety when using the lift Golvo 9000 mobile lift Instructions for Use Golvo 9000 Prod No 2000045 Golvo 9000 LowBase Prod No 2000049 7EN140108 Rev 4 ENGLISH 2020 QRG IMPORTANT Lifting and tr...

Page 2: ...ctions 4 Definitions 5 Technical Data 5 Dimensions 6 EMC Table 7 Assembly 9 Disassembly 12 Operation 12 Charging the Batteries 18 Maximum Load 19 Recommended Lifting Accessories 19 Recycling Instructi...

Page 3: ...line underneath the symbol indicate this product have been placed on the market after 2005 UL Recognized Component Mark for Canada and the United States EFUP Environmental Friendly Usage Period years...

Page 4: ...uestions please consult the responsible assistive device technician or the supplier The lift may not be used in areas where flammable mixtures may occur for example in areas where flammable goods are...

Page 5: ...mm 3 in twin wheels with brakes Turning diameter Golvo 9000 1330 mm 52 in Golvo 9000 LowBase 1330 mm 52 in Emergency lowering device Mechanical and electrical Lifting Speed no load 37 mm s 1 5 in s or...

Page 6: ...9000 LowBase B4 Model A B B1 B2 B3 B4 C D D2 E F F1 G Lmax Lmin M O max min max min max min Golvo 9000 2090 1455 1185 870 480 480 1020 735 907 623 0 1100 105 25 940 1816 571 436 391 Golvo 9000 LowBas...

Page 7: ...ge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humi...

Page 8: ...s the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequenc...

Page 9: ...on distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the t...

Page 10: ...ightening the two pre assembled M8 screws A B 8 Optional accessories Quick reference guide to be placed in Holder for Quick reference guide on the mast 7 Hang the hand control on the handle Motor for...

Page 11: ...single beep is heard service interval activated Alternatively use the push buttons simultaneously for emergency lifting up and down on the control box 11 Place the charger cable on the hook provided...

Page 12: ...e the lift using the push buttons on the handcontrol For raising and lowering Directional arrows show the direction of movement up down The lifting and base movement stops as soon as the push button i...

Page 13: ...rning check cables and connections Warning is shown until repaired Overload Too much load is applied to the lift Service needed contact Hill Rom SHORT CIRCUIT CHECK CONNECTION kg Ibs service 6 4 3 5 2...

Page 14: ...as follows Push down the red emergency lowering handle at the same time as loading the sling bar pull down the sling bar with your other hand This extends the strap and the sling bar is lowered Repea...

Page 15: ...d in the parking panel the lift should not be raised since this could be dangerous and cause personnel injury or damage to the lift if the sling bar should release and swing out from the panel Install...

Page 16: ...nd smooth and free from gravel debris ice and potholes Two caregivers are recommended for this task A gurney or a wheel chair should be in the immediate vicinity of the car when performing the lift Re...

Page 17: ...caregiver must guide the sling bar and patient out of the car and ensure that the head of the patient is guided safely while exiting the car The other caregiver will need to raise the lift while simul...

Page 18: ...box internal charger standard Plug the charger cable into mains 100 240 VAC see charger information 1 2 above The battery is fully charged after about 6 hours and the charger disconnects automaticall...

Page 19: ...e lift is affected Before changing lifting accessories you should always ensure that the lift after change can fulfil the desired lifting height in order to manage the lifting situations for which the...

Page 20: ...ormation LikoScale device for weighing a patient in combination with Golvo mobile lift LikoScale 350 Max 400 kg 880 lbs Prod No 3156228 LikoScale 350 is certified according to the European Directive N...

Page 21: ...nsfer and compose your lift system based on the patient need s and safety Read the User Manual for more information This Guide do not replace user manuals Mobile lift Maxload 200 kg Sling bar Maxload...

Page 22: ...tery capacity Check if the Li ION battery has been set in to sleep mode see chapter Operation 5 If the problem persists please contact Hill Rom The charger doesn t work 1 Check that the charger cables...

Page 23: ...own the entire lift starting from the top down Do not use a cloth that is dripping wet To have access to all areas run the lift into the highest and lowest positions and extend the base width adjustme...

Page 24: ...Golvo and ceiling lifts Alcohol Isopropyl alcohol 70 5 0 7 0 Viraguard Veridien Hand controls for all lifts Quaternary ammonium n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl ethyl...

Page 25: ...o product Expected Life Time The product has an expected service life of 10 years when correctly handled serviced and periodically inspected in accordance with Liko s instructions Parts listed below a...

Page 26: ...Copyright Liko R D AB 2020 ALL RIGHTS RESERVED Liko AB Nedre v gen 100 975 92 Lule Sweden 46 0 920 474700 www hillrom com Liko AB is a subsidiary of Hill Rom Holdings Inc...