Lighthouse ActiveCount100 Operating Manual
248083444-1 Rev 2
P a g e
8 of 40
Touch Screen Calibration
The ActiveCount100H 3.5” color touch screen is calibrated prior to shipping and there should be no need
to recalibrate at any time. Alternatively touch screen calibration may be performed at any time by
following the steps outlined here.
Touch Screen Calibration
1.
Power off the ActiveCount100H.
2.
While holding your finger on the touch screen; Power on ActiveCount100H and continue to hold
your finger on the touch screen until the “Performing touch screen calibration” screen appears.
3.
Follow the prompts to calibrate the touch screen by pressing and releasing on the filled circles
for all four corners.
Performing touch
Screen calibration.
Touch screen to
Continue.
Press & Release
On the filled
circle
Press & Release
On the filled
circle
To repeat calibration,
Restart the unit while
Pressing the screen
Until the calibration
Prompt appears
Touch screen to
Continue.
Press & Release
On the filled
circle
Press & Release
On the filled
circle