R5104V MODBUS Register Map v1.44
248083293-1 Rev 7
A-13
represent raw counts only.
Data Units Registers
The 33xxx register series identifies the units used by data items in the
30xxx series. The Units Registers run in parallel with the Data
Registers. For example, Data Register 30010’s Units Register is
33010.
Note:
Not all data types
have units.
Units are stored as 4 character ASCII strings across 2 registers. A
Units string containing less than 4 characters or no characters is padded
with NULLs but a Units string using all 4 characters does not end with
a NULL.
Table A-14 Examples of Particle Data Items
String
Description
0.3
Particle type of size 0.3 micron
1.0
Particle type of size 1.0 micron
20.0
Particle type of size 20.0 micron
.015
Particle type of size 0.015 micron or 15 nanometer
Summary of Contents for REMOTE 5104V
Page 1: ......
Page 2: ...00...
Page 3: ...Lighthouse Worldwide Solutions REMOTE 5104V Airborne Particle Counter Operating Manual...
Page 10: ...Lighthouse REMOTE 5104V Operating Manual t iv 248083293 1 Rev 7...
Page 12: ...Lighthouse REMOTE 5104V Operating Manual LOF 2 248083293 1 Rev 7...
Page 14: ...Lighthouse REMOTE 5104V Operating Manual LOT 2 248083293 1 Rev 7...
Page 16: ...Lighthouse REMOTE 5104V Operating Manual ii 248083293 1 Rev 7...
Page 24: ...Lighthouse REMOTE 5104V Operating Manual 2 4 248083293 1 Rev 7...
Page 46: ...Lighthouse REMOTE 5104V Operating Manual 4 6 248083293 1 Rev 7...
Page 72: ...Lighthouse REMOTE 5104V Operating Manual C 2 248083293 1 Rev 7...
Page 76: ...Lighthouse REMOTE 5104V Operating Manual I 4 248083293 1 Rev 7...
Page 77: ...00...
Page 78: ......