ApexRp Operators Manual
33
248083480-1 Rev 3
Figure 4-2 Serial COM Port
The connector pinouts are shown in Table 4-4.
RJ45 Pin
Signal Name
1
RS-232 TX (Output)
2
RS-232 RX (Input)
3
RESERVED for future use
4
RS-485B
5
RS-485A
6
RESERVED for future use
7
24 VDC
8
GROUND
Table 4-4 RJ45 Pinouts
To connect the instrument to an RS-485 network:
1. Make sure the SmartPort Cable is disconnected from the
instrument.
2. Install the ApexRp
in a perpendicular position with its Inlet
barb upward. Connect one end of a CAT5e cable to the Serial COM
port on the instrument (shown in Figure 4-2).
3. Connect the other end of the cable to an available RS485 port on
an LWS 485 Gateway
,
an LWS System Control Cabinet RS485 port or
other similar equipment port.