27
RF
emissions
CISPR 11
Group 1
The Infrared Thermometer LS-320
uses RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to cause
any interference in nearby electronic
equipment.
RF
emissions
CISPR 11
Class B
The Infrared Thermometer LS-320 is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer LS-320 is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Infrared Thermometer LS-320 should assure that it is used in such
an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
±6kV
contact
±8 kV
air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
3 A/m
3 A/m
Power frequency
magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital