23
● After removing a glucose test strip from its vial, immediately close
the vial cap tightly.
● Touch the blood glucose/β-ketone/total cholesterol test strips with
clean and dry hands.
● Use each test strip immediately after removing it from the vial.
● Write the opening date on the vial label when you first opened it.
Discard remaining blood glucose/β-ketone/total cholesterol test
strips after 6 months.
● Do not use test strips beyond their expiry date. This may cause
inaccurate results.
● Do not bend, cut, or alter a test strip in any way.
● Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.
Important Control Solution Information
● Use only Our control solution with your meter.
● Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
● It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
● Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
● Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do
NOT
freeze.