14
QUALITY CONTROL TESTING
When Should the Control Solution Test be
Performed?
● if it is mandatory following the local regulations in your country,
● if you suspect your meter or test strips are not working properly,
● if your blood glucose or ketone test results are not consistent with
how you feel, or if you think the results are not accurate,
● to practice the testing process, or
● if you have dropped or think you may have damaged your meter.
Test strips
(c)
, control solutions
(d)
, lancing device
(e)
or sterile
lancets
(f)
may not be included in the kit (please check the contents
on your product box). They can be purchased separately. Please
make sure you have those items needed for a test beforehand.
Performing a Control Solution Test
To perform a control solution test, you will need:
(b)
,
(c)
and
(d)
.
1. Insert the test strip to turn on your meter
Wait for the meter to display “
”, “ ”, and “
GLU
”, “
KETONE
” or
“
CHOL
”.
2. Press M to mark this test as a control solution test (for the
total cholesterol test)
With “QC” displayed, the meter will store your test result in memory
under “QC”. If you press
M
again, the “QC” will disappear and this test
1.
Start with the meter switched off. Insert a test strip to turn on the
meter. The screen will display “
”, a flashing “ ”, "
Gen
" and
“
CHOL
”.
2.
Press
M
to switch between General and QC mode.