WARNING: Do not spray the meter directly with solutions
as this could cause the solution to enter the case and
damage the electronic components.
WARNING: Do not immerse the meter or hold the meter
under running water.
WARNING: Cleaning and disinfection may in rare cases
damage the meter. Damage may include plastic housing
cracks, cloudiness or frosting of the LCD display, response
issues with the display, battery compartment fluid leakage, or
test port damage. Signs of meter performance deterioration
may include failure to recover proper control solution results
or the inability to perform a blood glucose test. If you observe
damage due to cleaning and disinfection, stop using the
meter and contact LifeScan Customer Service.
Cleaning and Disinfecting the Meter
Summary of Contents for OneTouch SS Xpress