9. Remove the test strip and discard into an appropriate
biohazard container.
NOTE: Acceptable control assay ranges are printed on
the Glucose Control Solutions vial label. If a QC test does
not fall within the specified range, verify that the
OneTouch
StatStrip Glucose Hospital Test Strips and OneTouch
StatStrip Glucose Control Solutions are not past their
expiration dates. Repeat the test with a new strip. If the
second test fails, inspect and clean the meter according to
the Cleaning and Disinfecting the Meter section. If the third
test fails, contact your local Customer Service representa-
tive, refer to www.onetouchstatstrip.com.
Testing a Quality Control Solution