Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
EN 60601-1-2
The T4 is intended for use in the electromagnetic environment specified below. The customer or
user of the T4 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emissions
CISPR11/EN55011
Group 1
The T4 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11/EN55011
Class B
The T4 is suitable for use in all establishments, including domestic es-
tablishments and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic pur-
poses.
Note: Only the recommended or supplied PC must be used in the sys-
tem to ensure compliance.
Harmonic emissions
EN 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions EN 61000-3-3
Complies
Electromagnetic Immunity
EN 60601-1-2
The T4 is intended for use in the electromagnetic environment specified below. The customer or
user of the T4 should assure that it is used in such an environment.
Immunity Test
EN 60601-1-2
Test Level
Compliance Level
Electromagnetic Environment Guid-
ance
Electrostatic dis-
charges (ESD)
EN 61000-4-2
+/- 6 kV:Contact
+/- 8 kV:Air
+/- 6 kV:Contact
+/- 8 kV:Air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at
least 30%
Electrical fast Transi-
ents/burst
EN 61000-4-4
Compliance is provided
by the recommended PC
equipment.
Compliance is provided
by the recommended PC
equipment.
Mains power should be that of a typical com-
mercial and/or hospital environment
Surge
EN 61000-4-5
Compliance is provided
by the recommended PC
equipment.
Compliance is provided
by the recommended PC
equipment.
Mains power should be that of a typical com-
mercial and/or hospital environment
Voltage dips,short in-
terruptions and volt-
age variations on
power supply input
lines
EN 61000-4-11
Compliance is provided
by the recommended PC
equipment.
Compliance is provided
by the recommended PC
equipment.
Mains power should be that of a typical com-
mercial and/or hospital environment. If the
user of the T4 requires continued operation
during power mains interruptions, it is rec-
ommended that the T4 be powered from an
uninterruptible power supply or a battery
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial and/or hospital en-
vironment