T4 EEG Amplifier User Manual
25
Figure 12 Recording Channel editing
4.2
Configuring the amplifier
When you have finished setting up the recording protocol, connect the T4 to the host computer.
Steps 5 to 9 describe configuration and set-up of an EEG recording.
Step 5 Activate the recording control
The Trackit application software will normally connect automatically to the T4 Amplifier as soon as
it is plugged in. Otherwise, from the Trackit toolbar select the Trackit Control Panel (‘handshake’)
icon which will allow for manual connection.
Figure 13 Trackit software toolbar
Key
1 New Patient
2 Setup Recording
3 Trackit Control Panel
4 Ongoings On
5 Ongoings Off
6 Impedance Check On
7 Calibration On
8 Calibration Off
9 Page Down
10 Page Up
11 Get Trackit Events
12 Email Events List
13 Notch Filter On/Off
14 PC Record On
15 PC Record Off
16 Videometry (optional)
17 Photic Stimulation
18 Hyperventilation
19 Vertical sensitivity
20 Chart speed
21 Open files for playback
22 Stop playback
23 Start playback
24 Page back
25 Stop paging
26 Page forward
27 Paging speed
28 Playback time
29 Montage selection
30 Lo-filter selection
31 Hi-filter selection
Clicking on the ‘handshake’ icon brings up the Control Panel dialog box.