58
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The BM 66/U is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 66/U should assure
that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V
rms
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of
the BM 66/U, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d
= 1.2
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is a
ff
ected by absorption and reflection from struc-
tures, objects and people.