Page 1: ...experienced technician 30 minutes to 1 hour If the instru ment has been running recently the bench may be hot to the touch Let it cool for 5 or 10 minutes after powering off if needed Note Be sure that you are properly grounded to avoid elec trostatic discharge that can damage the electronics Use an anti static wrist strap electrostatic discharge grounding mat or occasionally touch bare metal that...

Page 2: ... Make sure that the foam insulation on the inside top cover is positioned over the optical bench it is required for thermal stability Reconnect the top cover ribbon cable and then install the top cover 10 In the LI 830 LI 850 and LI 870 reset the source hours to 0 The LI 830 LI 850 and LI 870 record the number or oper ating hours for the optical source You should reset this count after replacing t...

Page 3: ...ations you will measure A dual span however will bring the instrument performance within specifications across the full measurement range If you do not have zero and span gases send the instrument back to LI COR for a factory recalibration LI 8100 A Source and Detector Replacement The LI 8100A Soil Gas Flux System the source and detector on this optical bench can be replaced if they fail These ins...

Page 4: ...of the white panel 7 Lift the assembly out of the yellow case and turn it upside down 8 Remove the 6 screws from the optical bench cover 9 Leaving the hoses connected to the optical bench lift the bench out of the foam casing 10 There are four screws on the source and detector housings that must be removed Remove only the screws from the assembly you are repla cing Remove the four screws in the co...

Page 5: ... of the bench Note that the orientation of the optical bench cyl inder is not important either end can be inserted into the source or detector housing 15 Replace the tubing Any tubing that you removed should be replaced with a piece of equal length which can be cut from the tubing that is included in the instrument spares kit 16 Re assemble the LI 8100 A case Make sure that the foam insulation on ...

Page 6: ...6 ...

Page 7: ...7 ...

Page 8: ...6 U S and Canada envsales licor com LI COR Distributor Network www licor com envdistributors Regional Offices LI COR Biosciences GmbH Siemensstraße 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 envsales gmbh licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambridge CB4 0WS United Kingdom Phone 44 0 1223 422102 envsales UK licor com ...