background image

On the LI-840/A, there are four spacers between the shield
and detector. Collect the four spacers.

3

Separate the detector from the optical path.

Be sure that the inside of the optical path is clean. It
should be free of dust and other contaminants. Clean it if
needed, as described in the instruction manual.

4

Install the new O-ring that is included with the
detector.

5

Install the replacement optical detector.

On the LI-840/A, place the four spacers on the screws so
they fit between the shield and detector circuit board.
Tighten the screws firmly.

6

Press the tube over the hosebarb—be sure it covers
both barbs on the hosebarb.

7

Work the optical bench into the foam.

8

Connect the wiring harnesses to the detector and
source.

9

Reassemble the instrument.

Make sure that the foam insulation on the inside top
cover is positioned over the optical bench; it is required
for thermal stability. Reconnect the top cover ribbon
cable and then install the top cover.

Important:

LI-840's that are running older software must

be updated to the latest LI-840A embedded software,
which is available for free from

licor.com/env/support/LI-

840A/software.html.

10

Zero and span the analyzer as described in the instruc-
tion manual.

A simple zero and span will be adequate if the span gas is
close to the concentrations you will measure. A dual span,
however, will bring the instrument performance within
specifications across the full measurement range. If you
do not have zero and span gases, send the instrument
back to LI-COR for a factory recalibration.

LI-8100/A Source and Detector
Replacement

The LI-8100A Soil Gas Flux System the source and detector
on this optical bench can be replaced if they fail. These
instructions describe how to change the source, detector
and/or optical bench. A Phillips head screwdriver and a 1/2”
open end wrench are the only tools needed to replace the
source and/or detector in all instruments. To replace the
source or detector:

Component

Part Number

Source

8100-902

Detector

800-906

Optical Bench

800-904

Note:

Be sure that you are properly grounded to avoid elec-

trostatic discharge that can damage the electronics. Use an
anti-static wrist strap, electrostatic discharge grounding
mat, or occasionally touch bare metal that has a clear path
to ground, such as an unpainted computer case.

1

Power the instrument off and remove the battery or
other external power source.

Allow the instrument optical bench to cool for about 10
minutes.

2

Open the Analyzer Control Unit and remove the
access panel by loosening the 4 thumbscrews.

3

Remove both air filters by pressing the orange part of
the quick-connect fittings toward the white part of
the connector and pulling the filter out.

Discard the filters. There are replacement filters included
that can be used during reassembly.

3

Summary of Contents for LI-8 0 Series

Page 1: ...experienced technician 30 minutes to 1 hour If the instru ment has been running recently the bench may be hot to the touch Let it cool for 5 or 10 minutes after powering off if needed Note Be sure that you are properly grounded to avoid elec trostatic discharge that can damage the electronics Use an anti static wrist strap electrostatic discharge grounding mat or occasionally touch bare metal that...

Page 2: ... Make sure that the foam insulation on the inside top cover is positioned over the optical bench it is required for thermal stability Reconnect the top cover ribbon cable and then install the top cover 10 In the LI 830 LI 850 and LI 870 reset the source hours to 0 The LI 830 LI 850 and LI 870 record the number or oper ating hours for the optical source You should reset this count after replacing t...

Page 3: ...ations you will measure A dual span however will bring the instrument performance within specifications across the full measurement range If you do not have zero and span gases send the instrument back to LI COR for a factory recalibration LI 8100 A Source and Detector Replacement The LI 8100A Soil Gas Flux System the source and detector on this optical bench can be replaced if they fail These ins...

Page 4: ...of the white panel 7 Lift the assembly out of the yellow case and turn it upside down 8 Remove the 6 screws from the optical bench cover 9 Leaving the hoses connected to the optical bench lift the bench out of the foam casing 10 There are four screws on the source and detector housings that must be removed Remove only the screws from the assembly you are repla cing Remove the four screws in the co...

Page 5: ... of the bench Note that the orientation of the optical bench cyl inder is not important either end can be inserted into the source or detector housing 15 Replace the tubing Any tubing that you removed should be replaced with a piece of equal length which can be cut from the tubing that is included in the instrument spares kit 16 Re assemble the LI 8100 A case Make sure that the foam insulation on ...

Page 6: ...6 ...

Page 7: ...7 ...

Page 8: ...6 U S and Canada envsales licor com LI COR Distributor Network www licor com envdistributors Regional Offices LI COR Biosciences GmbH Siemensstraße 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 envsales gmbh licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambridge CB4 0WS United Kingdom Phone 44 0 1223 422102 envsales UK licor com ...

Reviews: