background image

3

Separate the source from the optical path.

Be sure that the inside of the optical path is clean—free of
dust and other contaminants. Clean it if needed, as
described in the instruction manual.

4

Install the new O-ring that is included with the
source.

5

Install the replacement optical source.

Tighten the screws firmly.

6

Press the air-in tube over the hosebarb—be sure it
covers both barbs on the hosebarb.

Removing a section of tubing will reduce the total
volume of the analyzer, and this may affect some char-
acteristics of the analyzer, such as the T90 specification,
but the reduction in volume is generally negligible. There
is enough slack in the original tubing for a single replace-
ment of the source, but if you replace it more than once,
you may need to replace the entire segment of tubing.
Should you choose to recalculate system volume after
removing a segment of tubing (which is not required), the
Bev-A-Line® has an inside diameter of 1/8" (0.3175 cm).
If you removed 1 cm (0.39 inches) of tube, the volume
removed is about 0.079 cm

3

(π × r

2

× length; 3.14 ×

(0.3175 cm / 2)

2

× 1 cm = 0.079 cm

3

).

7

Work the optical bench into the foam.

Be sure that the bench is secured tightly in the foam insu-
lation and that no parts are in direct contact with the sur-
rounding metal cradle.

8

Connect the ribbon cables that were removed pre-
viously.

9

Reassemble the instrument.

Make sure that the foam insulation on the inside top
cover is positioned over the optical bench; it is required
for thermal stability. Reconnect the top cover ribbon
cable and then install the top cover.

10

In the LI-830, LI-850, and LI-870, reset the source
hours to "0".

The LI-830, LI-850, and LI-870 record the number or oper-
ating hours for the optical source. You should reset this
count after replacing the source. Connect using the soft-
ware and then click

Settings > Diagnostics

and click the

gear button near

Source

, and then reset the hours.

11

Zero and span the analyzer as described in the instruc-
tion manual.

A simple zero and span will be adequate if the span gas is
close to the concentrations you will measure. A dual span,
however, will bring the instrument performance within
specifications across the full measurement range. If you
do not have zero and span gases, send the instrument
back to LI-COR for a factory recalibration.

Replacing the Optical Detector

You can also replace the optical detector.

1

Remove the tube from the detector.

When replacing the detector on the LI-830, LI-850, or LI-
870, you will probably need to replace the entire segment
of urethane tubing, because cutting a portion to remove it
from the hose barb will not leave enough tubing to re-con-
nect using the original segment. A replacement segment
of urethane tubing is included in the detector replace-
ment kit for this purpose. On the LI-800, LI-820, or LI-
840/A, the tubing is long enough to reconnect.

2

Remove the four screws that secure the optical
detector and shield.

2

Summary of Contents for LI-8 0 Series

Page 1: ...experienced technician 30 minutes to 1 hour If the instru ment has been running recently the bench may be hot to the touch Let it cool for 5 or 10 minutes after powering off if needed Note Be sure that you are properly grounded to avoid elec trostatic discharge that can damage the electronics Use an anti static wrist strap electrostatic discharge grounding mat or occasionally touch bare metal that...

Page 2: ... Make sure that the foam insulation on the inside top cover is positioned over the optical bench it is required for thermal stability Reconnect the top cover ribbon cable and then install the top cover 10 In the LI 830 LI 850 and LI 870 reset the source hours to 0 The LI 830 LI 850 and LI 870 record the number or oper ating hours for the optical source You should reset this count after replacing t...

Page 3: ...ations you will measure A dual span however will bring the instrument performance within specifications across the full measurement range If you do not have zero and span gases send the instrument back to LI COR for a factory recalibration LI 8100 A Source and Detector Replacement The LI 8100A Soil Gas Flux System the source and detector on this optical bench can be replaced if they fail These ins...

Page 4: ...of the white panel 7 Lift the assembly out of the yellow case and turn it upside down 8 Remove the 6 screws from the optical bench cover 9 Leaving the hoses connected to the optical bench lift the bench out of the foam casing 10 There are four screws on the source and detector housings that must be removed Remove only the screws from the assembly you are repla cing Remove the four screws in the co...

Page 5: ... of the bench Note that the orientation of the optical bench cyl inder is not important either end can be inserted into the source or detector housing 15 Replace the tubing Any tubing that you removed should be replaced with a piece of equal length which can be cut from the tubing that is included in the instrument spares kit 16 Re assemble the LI 8100 A case Make sure that the foam insulation on ...

Page 6: ...6 ...

Page 7: ...7 ...

Page 8: ...6 U S and Canada envsales licor com LI COR Distributor Network www licor com envdistributors Regional Offices LI COR Biosciences GmbH Siemensstraße 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 envsales gmbh licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambridge CB4 0WS United Kingdom Phone 44 0 1223 422102 envsales UK licor com ...

Reviews: