Making Measurements
Matching the Analyzers
Using the LI-6400
4-37
in match mode, or not at all. Pressing
f1
(
exit
) will cause the match valve to
toggle back to normal position, and the exit countdown will commence.
Messages in Match Mode
Match mode has some messages for alerting you to possible problems:
"
CO2R DidnÕt Change
Ó
(OPEN 3.2) After the H2O reference reading has stabilized, if the message
appears, it is because the CO
2
reference reading changed less than
1.5 µmol mol
-1
after the match valve closed, and the expected change was
much larger than that. Reasons for this would be a match valve that is stick-
ing, or the air flow tube connecting the chamber to the match valve not being
in place, or some other flow related problem.
IRGA Match Mode
ÒCONFIG.JoeÓ
CO2R_µml CO2S_µml H2OR_mml H2OS_mml
330.1 329.2 18.24 18.55
Prev: 00:22:01 CO2=402 H2O=18.8
exit
MATCH
Elapsed time
Previous values
Figure 4-6. The display while in Match Mode. The elapsed time is the time since the
previous match, and the previous values are the concentrations at which the last
match occurred.
IRGAs
CO2R didn't change enough. Match
valve OK? Return tube in place?
Warning
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...