Guidance and- manufacturer's declaration.
Electromagnetic immunity
The device is intended for use in. the electromagnetic
environment specified below. The customer or the user
should assure that it is used in such an environment.
Immunity
test
IEC 60501
test level
Compliance
level
Conducted
RF lEC
61000-4-6
3 Vrms
150 kHz
to 80 MHz
3 Vrms
3 V/m
Electromagnetic
environment -
guidance
, 80MHz
to 800MHz
, 800MHz
to 2,5MHz
Portable and mobile RF
Communications
equipment should be used
no closer to any part of
the device, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
Radiated
RF lEC
61000-4-3
3 V/m80
MHz to
2.5 GHz
35
d
=[ ]
p
3.5
V1
d
=[ ]
p
3.5
E1
d
=[ ]
p
7
E1