Installation
39
300554859_002_C0 - 11/2016 - © Leybold
Purge gas connection with purge gas valve
Unscrew and remove the locking screw and the gasket from the purge gas
connection of the TURBOVAC.
Screw in the purge gas valve and the gasket into the thread, then plug in the
corresponding cable into the accessories connection X204. Thereafter
change the function for the accessory connection to "Always On" or "Active
at Start Command", as the default is always set to Off. To change the venting
valve function code go to Menu
→
Accessory
→
Purge (via the TPU inter-
face).
Connect the purge gas supply at the valve’s inlet (G1/8“).
Purge gas connection with purge gas throttle
Unscrew and remove the locking screw and the gasket from the purge gas
connection of the pump.
Then screw in the throttle and the gasket into the thread.
Fig. 3.13 Fitting valve
Ventil
800120V0012
800120V0013
800120V0022
Sperrgasdrossel G1/8"
800120V0014
Luftfilter
800110V0022
Valve
800120V0012
800120V0013
800120V0022
Air filter
800110V0022
Purge gas throttle G1/8"
800120V0014