Operation
68
300554859_002_C0 - 11/2016 - © Leybold
Fig. 4.10 Configuration menu for the gauges
Gauges
Here you may select between the units of measurement millibar, Torr and
Pascal via the drop-down list. These settings are carried over to the start
screen. The types of vacuum gauge, the values currently measured and the
vacuum gauge head status are displayed. Through OFF the vacuum gauge
head may be switched off. The gas type correction factor needs to be set up
according to the Operating Instructions for the respective vacuum gauge
head. You may enter the value through the entry field or the slider.
Turbomolecular pump (Menu TMP)
Here you may select between two options:
■
The turbomolecular pump starts upon issuing the start command. Here the
entered pressure values are disregarded.
■
The turbomolecular pump starts depending on the pressure as soon as the
pressure as measured by the selected vacuum gauge head has dropped
below the preset threshold. The threshold may be entered directly or set
up through a slider.
Parameters
Through the parameter menu you may select via the drop-down list the
respective parameters. When wanting to change a parameter you may do
this through the entry field or through the slider. Through SET and SAVE, the
values are saved.
System Menu
User Profile
Through the user profile you may change the log-in data for the web server.