Operation
54
GA10218_002_C0 - 10/2016 - © Leybold GmbH
3.6.2 View
Main menu > View
In this menu Fig.24 all features that influence the data display are put together.
3.6.3 Scale Linear/Logarithmic
Main menu > View > Scale linear/logarithmic
These settings apply to the bargraph (= bar underneath the digital figures in the
measurement mode) and Y-axis in the trend diagram.
The scale of the bargraph can either be linear or logarithmic. With the arrows
(
and
) it can be determined how many decades the bargraph and Y-axis are
covered. Usually a logarithmic scale is recommended because leak rates may
change easily over several decades. Default setting is logarithmic with 4 decades.
Fig. 24 Display: View Menu
Explanation to Fig.24:
Key No. Name
Description
1
Back
Return to the main menu.
2
Scale linear/logarithmic Settings for bargraph and trend mode.
Please refer to Chapter
3
Display range auto/
manual
Manual or automatic scaling. Please
refer to Chapter
4
Time axis
Time axis in trend. Please refer to
Chapter
5
Contrast
Display contrast. Please refer to Chap-
ter
6
Background in standby Background displayed or not. Please
refer to Chapter
8
Lower display limit
Setting of the display limit. Please refer
to Chapter
Summary of Contents for 250000V01
Page 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Page 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Page 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Page 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Page 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Page 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Page 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...