1.3 Technical Data
1.3.1 Technical Data PHOENIX L300i
Physical data
To get down to the minimum detected leak rate range some conditions must be
fulfilled:
PHOENIX L300i has to run at least 20 minutes
Ambient conditions must be stable (temperature, no vibration/accelerations.)
The part under test has been evacuated long enough without using the zero func-
tion (background is no longer decreasing)
ZERO must be active
Max. inlet pressure
15 mbar
Minimum detectable Helium leak rates
- in vacuum mode
<5·10
-12
mbar l / s
- in sniffer mode
<1·10
-7
mbar l / s
Minimum detectable Hydrogen leak rates
- in vacuum mode
<1·10
-8
mbar l / s
- in sniffer mode
<1·10
-7
mbar l / s
Maximum Helium leak rate which can be displayed 0.1 mbar l / s
Measurement range
12 decades
Time constant of the leak rate signal (blanked off,
63% of the final value)
<1 s
Pumping speed (Helium) at the inlet
- GROSS mode
0.4 l/s
- FINE mode
> 2.5 l/s
Detectable masses
2, 3 and 4
Mass spectrometer
180° magnetic sector
field
Ion source
2 filaments;
Iridium/Yttrium-oxide
Inlet port
DN 25 KF
Run-up time (after starting)
< 2 min
Physical data
Summary of Contents for 250000V01
Page 104: ...Waste Disposal GA10218_002_C0 10 2016 Leybold GmbH 104...
Page 105: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 10 7 Certificates Fig 43...
Page 106: ...Certificates GA10218_002_C0 10 2016 Leybold GmbH 106...
Page 108: ...Appendix 107 GA10218_002_C0 10 2016 Leybold Vacuum...
Page 109: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum 108...
Page 110: ...Appendix 109 GA10218_002_C0 10 2016 Leybold Vacuum...
Page 111: ...Appendix GA10218_002_C0 10 2016 Leybold Vacuum110...