Troubleshooting
Symptom
Possible Cause/Remedy
No digital communication
-
Check electrical connections (3 wires from transmitter to
communication equipment)
-
Transmitter and communication equipment baud rates have
to match
-
Use of incorrect transmitter address. Try address 254
-
Attention characters missing (@)
-
Termination characters missing (;FF)
NAK180 is received when
transmitting setpoint commands
-
The transmitter setup is locked. Refer to disable lock
procedure in the communication protocol.
Incorrect pressure value
-
Other gas present than transmitter gas setting or trace of
gas.
-
Contaminated sensor. Transmitter repair required.
-
Corroded sensor. Transmitter repair required.
-
Perform a zero adjustment/FS adjustment.
-
Check Setting of Controller or display.
Incorrect pressure value at low
pressure
-
Contaminated sensor. Transmitter repair required.
-
Corroded sensor. Transmitter repair required.
-
Incorrect zero adjustment has been executed.
-
Transmitter exposed to heat or cooling air stream.
-
Perform a zero adjustment.
Incorrect pressure value at high
pressure.
-
Contaminated sensor. Transmitter repair required.
-
Corroded sensor. Transmitter repair required.
-
Incorrect FS adjustment has been executed.
-
Other gas or gas trace present than transmitter gas setting.
-
Perform a FS adjustment.
Cold cathode does not provide
measurements
-
If the cold cathode is turned on at low pressure <1.00E-7
mbar or is contaminated a delay can occur from turning on
high voltage to the ionization begins. At low pressure the
delay can be several minutes.
Cold cathode value is lower than
actual pressure
-
If the cold cathode sensor has been operated a too high
pressure or exposed to fast pressure cycles the sensor can
be sputtered. Sputtering of sensor will normally provide
lower reading than actual pressure.
Setpoint relay does not trip
-
Setpoint not enabled.
-
Setpoint value not set to proper value.
-
Setpoint direction is different than the user expects.
-
Check electrical connection.
-
Check part number to see if transmitter has setpoint relays.
No analog output
-
Power supply turned off.
-
Check electrical connections.
Status LED illuminating red
-
Sensor element defect.
Summary of Contents for 230070V02
Page 10: ...10 300544652_002_C1 11 2016 Leybold P N 230085V02 P N 230089V02 Flanges KF 25 KF 40 CF 40...
Page 35: ...35 300544652_002_C1 11 2016 Leybold...
Page 36: ...36 300544657_002 10 2015 Leybold Declaration of Conformity...
Page 37: ...37 300544657_002 10 2015 Leybold Notes...
Page 38: ...38 300544657_002 10 2015 Leybold Sales and Service...