![LEYBOLD 230070V02 Operating Manual Download Page 15](http://html1.mh-extra.com/html/leybold/230070v02/230070v02_operating-manual_1890893015.webp)
15
300544652_002_C1 - 11/2016 - © Leybold
4.1
Input/Output Wiring
To comply with EN61326-1 immunity requirements, use a braided, shielded cable. Connect the braid to the metal
hoods at both ends of the cable with the end for power supply connected to earth ground.
Part Numbers: 230070V02, 230071V02
,
230072V02, 230085V02:
I/O Connector (8 pin RJ45/FCC68)
PIN Description
1
Power + (9-30 VDC)
5
Analog Output -
2
Power return - (GND)
6
Status
3
Analog
7
Not Connected
4
ID-Resistor, 85
kΩ (±1%)
8
Not Connected
Part Numbers: 23088V02:
I/O Connector (15 pin HD Sub D male )
PIN Description
1
RS232 Transmit
9
Relay 1, Normally Closed
2
RS232 Receive
10
Relay 2, Normally Closed
3
Power + (9-30VDC)
11
Relay 2, Common
4
Power return - (GND)
12
Relay 2, Normally Open
5
Analog
13
Relay 3, Normally Closed
6
Analog Output -
14
Relay 3, Common
7
Relay 1, Normally Open
15
Relay 3, Normally Open
8
Relay 1, Common