Page 1: ...Oil free Vacuum Pumps EcoDry M 15 130 000 130 001 130 002 130 004 EcoDry M 20 130 010 130 011 130 012 130 014 Operating Instructions GA 01 701 5 02 Vacuum Solutions Application Support Service LEYBOL...

Page 2: ...ormity 16 Figures The references to figures e g 1 2 consist of the Fig No and the Item No in that order Leybold Service If a pump is returned to Leybold indicate whether the pump is free of substances...

Page 3: ...M 20 with three phase motor Cat No 130 010 with single phase motor 115 V 60 Hz USA Cat No 130 011 with single phase motor 230 V 50 60 Hz Euro Cat No 130 012 with single phase motor 200 V 50 60 Hz Cat...

Page 4: ...gram The EcoDry M is not suited for pumping of combustible and explosive gases or vapours radioactive and toxic substances pyrophorous substances The EcoDry M must be integrated in the system control...

Page 5: ...sure that the lines exhaust and intake lines and other vacuum connections are clean and free of oil This applies in particular if oil sealed pumps were used before The conditions must be checked befor...

Page 6: ...The static seals made of Viton and the PTFE radial shaft seals which are lubri cated with a grease which is free of hydrocarbons and which are highly media compatible reliably prevent the exchange of...

Page 7: ...er requirements at inlet pressures 20 mbar 15 Torr W 300 300 Motor Mains voltage frequency 3 ph 50 Hz V 220 240 346 415 10 to 5 220 240 346 415 10 to 5 3 ph 60 Hz V 200 277 380 480 10 to 5 200 277 380...

Page 8: ...or 3 Phase motor 176 229 212 212 53 37 76 289 76 289 89 191 6 2 Exhaust silencer Option Vibration absorber Option 19 5 17 2 350 335 117 52 38 7 186 314 20 2 Fig 1 Dimensional drawing of the EcoDry M w...

Page 9: ...Hz M 15 60 Hz M 20 60 Hz M 20 50 Hz 100 1000 Inlet pressure Pumping speed Fig 2 Pumping speed vs inlet pressure for the EcoDry M 15 M 20 50 60 Hz operation Fig 3 Connection for the intake and exhaust...

Page 10: ...installed resp provided by the customer 2 2 Installation The EcoDry M pump is installed upright or horizontally on the anti vibration mounts provided Make sure that there is a clearance of at least 1...

Page 11: ...0 mbar abs 2 4 Electrical Connection Make sure that the mains supply is turned off when cables are connected The electrical connection should only be performed by staff qualified in accordance with VD...

Page 12: ...ctions must only be provided by an electrician in consideration of the requirements in accordance with DIN EN 60204 Part 1 The motor must be connected to the mains power as indicated in the drawing Be...

Page 13: ...ral Application The EcoDry M is ideal for applications that require the achievement of pressures as low as 2 10 2 mbar 2 Pa 15 mTorr without the risk of contamination from liquid lubricants The pump m...

Page 14: ...red If in spite of this any unexpected maintenance becomes necessary within this period please get in touch with the local Leybold Service When returning equipment to our Service please fill in comple...

Page 15: ...pipe Pump temperature too low Warm up the pump 1h Pumping speed of Dirt trap in the intake line is clogged Clean the dirt trap the pump is too Connection lines too narrow or too long Use sufficiently...

Page 16: ...y M 20 130 010 130 011 130 012 130 014 EC Declaration of Conformity The product meets the requirements of the following directives EC Directive on Machinery 98 37 EG EC Directive on Low Voltages 73 23...