GA01201_1002 - 07/2005
15
Installation
3
Installation
3.1
Placement
The standard pump is
not
suited for installation in the explosion hazard
areas. When planning such an application please contact us first.
The TRIVAC B pump can be set up on a flat, horizontal surface. Rubber
feet under the coupling housing ensure that the pump can not slip.
If you wish firmly install the pump in place, insert bolts through bore holes
in the rubber feet.
Max. tilt for the pump (without further attachment) with possibly fitted
standard accessories is 10° from the vertical.
The rubber feet act as vibration absorbers. They must therefore not be
compressed by screws.
When installing the TRIVAC B pump, make sure that the connections and
controls are readily accessible.
The site chosen should allow adequate air circulation to cool the pump
(keep front and rear unobstructed).
The ambient temperature should not 40 °C (104 °F) and not
drop below +12 °C (55 °F) (see Section 4.2.3).
The max. amount of heat given off approximately corresponds to the
rated motor power.
Fig. 7 Connections and controls
Warning
Caution
Key to Fig. 7
1
Eye for moving the pump
2
Intake port
3
Exhaust port
4
Oil-level glass
5
Gas ballast knob
6
Threaded connection
M 16 x 1.5 for inert gas ballast
7
Adapter
8
Cover plate
9
Cover plate; connection for oil filter