GA01201_1002 - 07/2005
1.4
Technical Data
We can only guarantee that the pump will meet its specifications when
using the type of lubricant which has been specified by us.
TRIVAC D 4 B
TRIVAC D 8 B
two-stage
two-stage
Nominal pumping speed
1)
m
3
· h
-1
4.8
9.7
Pumping speed
1)
m
3
· h
-1
4.2
8.5
Ultimate partial pressure without gas ballast
1)
mbar
10
-4
Ultimate total pressure without gas ballast
1)
mbar
< 2 · 10
-3
Ultimate total pressure with gas ballast
1)
mbar
< 5 · 10
-3
Water vapor tolerance
1)
mbar
30
25
Water vapor capacity
gm · h
-1
93
157
Oil filling, min./max.
l
0.3 / 0.8
0.3 / 0.9
Noise level * to DIN 45 635,
dB (A)
50 - 52
(without/with gas ballast)
Admissible ambient temperature
°C
12 - 40
Motor rating *
W
370
Nominal speed *
min
-1
1500
Type of protection
IP
54
Weight *
kg
18.7
21.2
Connections, intake and exhaust side
DN
16 KF
1)To DIN 28 400 and subsequent numbers
*)Lenght, weight, motor rating, nominal speed and noise levels for the pumps with AC motor, 230 V, 50 Hz, only.
*)a mains cable must be ordered additionally:
■
100 V - 120 V (NEMA)
Cat.-No.: 200 81 090
■
200 V - 240 V (NEMA)
Cat.-No.: 200 81 141
■
200 V - 240 V (EURO)
Cat.-No.: 200 81 091
Note
13
Description
1.4.1
Motor Related Data
Cat. No.
l
W (kg) Cat. No.
l
W (kg)
Motor connections
Motor
Rated
Speed
P/N
(mm)
(mm)
voltage, frequency
power
current
motor
113 07
223
10.6
113 17
248
12.8
ohne Motor
-
-
-
-
140 081
455
19.1
140 082
480
21.1
100 V-120 V, 50 Hz/60 Hz*
7.7/5.6 A
200 V-240 V, 50 Hz/60 Hz*
4.0/2.8 A
D 4 B D 8 B
1~
570 W/660 W
1420/1690
10000 2292