Installation
Warning
Turn off power
Check voltage
Warning
Warning
Warning
2.4
Electrical Connections
Wiring must be done by a qualified electrician. Otherwise, electric
shock or fire may occur.
Turn off incoming main electricical power before wiring the pump.
Failure to do so can cause injury from to electric shock.
Check eIectric power and voltage before doing the wiring.
Three-phase
motor has dual voltage 200/380V 50 Hz, 208/460 V 60 Hz.
Single-phase
motor has dual voltage 100/200V 50 Hz, 115/208-230V 60
Hz.
Check electric source and voltage, and wiring.
2.4.1
Single-phase motors
Be sure to install a reliable main disconnect swtich (or breaker) for
emergency stop. In addition, shut off the SC 5 D by turning off the
switch on the motor. Failure to do so can cause damage or fire.
The single phase motor has an internal thermal protector. The
motor restarts without warning after the protector trips.
CE Requirement
Min. circuit current capacity of conductor is 10 A. Max. branch circuit brea
ker is 15 A (industrial rated).
Protect the motor by installing the recommended breaker. (refer to
chart 1). Failure to do so can cause electrical shock or fire.
18
GA01423_0102 - 11/2004
Summary of Contents for SC 15 D
Page 1: ......
Page 7: ...Description SC 15 D SC 30 D Fig 3 Connections and controls SC 15 D 30 D GA01423_0102 11 2004 7...
Page 13: ...Description SC 15 D Fig 7 Dimensional drawing for the SC 15 D GA01423_0102 11 2004 13...
Page 14: ...Description SC 30 D Fig 8 Dimensional drawing for the SC 30 D 14 GA01423_0102 11 2004...
Page 30: ...Operation SC 15 D SC 30 D SC 5 D Fig 24 Position of Air Flush Port 30 GA01423_0102 11 2004...