17
ENGLISH
20.2 Manufacturer's guide and declaration – electromagnetic immunity
The Piuma UP system is intended to be used in the following electromagnetic conditions. The
client or user of the Piuma UP system must ensure that it is used in these conditions.
Immunity test
Test level IEC60601
CONFORMITY level
ELECTROMAGNETIC environment - guide
Electrostatic
discharge (ESD)
IEC 61000-4-2
6kV contact
±
8kV air
±
6kV contact
±
8kV air
The floor must be wood, concrete
or tile. If the floor is covered with
a synthetic material, the relative
humidity must be at least 30%
Electrical fast
transient/
burst immunity
IEC 61000-4-4
±
2kV for power
lines
±
1kV for I/O lines
±
2kV for power
lines
±
1kV for I/O lines
The quality of the mains voltage
must be that of a typical commercial
or hospital environment.
Surge voltage
IEC61000-4-5
±
1 kV line to line
±
2 kV line to earth
±
1 kV line to line
The mains power must be that
of a typical commercial or hospital
environment.
Immunity test
Test level IEC60601
CONFORMITY level
ELECTROMAGNETIC environment - guide
Voltage drop outs,
brief interruptions
and variations
of voltage on the
power supply lines
IEC61000-4-11
<5% UT
(>95% UT drop
out) for 0.5 cycles
40% UT
(60% UT drop out)
for 5 cycles
70 % UT
(30 % UT drop out)
for 25 cycles
<5% UT
(>95% UT drop
out) for 5 seconds
<5% UT
(>95% UT drop
out) for 0.5 cycles
40% UT
(60% UT drop out)
for 5 cycles
70 % UT
(30 % UT drop out)
for 25 cycles
<5% UT
(>95% UT drop
out) for 5 seconds
The quality of the mains supply
must be that of a typical commercial
or hospital environment. If the user
of the Piuma UP system requires
continuous functioning, even during
a power failure, it is recommended
to supply power to the Piuma UP
system with a UPS or with batteries.
Mains frequency
magnetic field
(50/60Hz)
IEC61000-4-8
3 A/m
3 A/m
Magnetic fields at the mains
frequency should have the typical
characteristics of a commercial
or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
The Piuma UP system is intended to be used in the following electromagnetic conditions. The
client or user of the Piuma UP system must ensure that it is used in these conditions.