IT 1911i / IT 1981i - 03
2015/09
Resetting the IT 1911i / IT 1981i to factory settings
To reset all parameters to factory settings, scan the adjacent
bar code.
Attention!
All settings are lost!!!
Return the hand-held scanner to the base station to apply the settings. This procedure is concluded with audible confirmation
signals.
You may then continue making settings or operation of the device.
Trigger
To activate the read process, a trigger signal is to be sent via the serial RS 232 interface or USB interface (COM port emulation
only). The command is to be sent at the set baud rate, parity, and data and stop bits.
The command for activation is:
SYN T CR
ASCII decimal values: 022; 084; 013
To cancel read readiness, send a deactivation.
The command for deactivation is:
SYN U CR
ASCII decimal values: 022; 085; 013
Following a successful read operation, the hand-held scanner deactivates itself.
The second option is activation via the built-in trigger button.
Configuration for the Leuze standard protocol
Scan the adjacent 2D-code.
The hand-held scanner is set to the following transmission parameters:
RS 232 transmission with 9,600 baud, 8 data bits, 1 stop bit, no parity, prefix <STX>,
postfixes <CR><LF>.
Return the hand-held scanner to the base station to apply the settings. This procedure
is concluded with audible confirmation signals.
Notice!
To charge the hand-held scanner, the power supply unit must be plugged in
and the hand-held scanner placed in the base station.
IT 1911i / IT 1981i