LEM ITZ 5000-S FLEX ULTRASTAB Quick Start Manual Download Page 13

Page 13/15 

28October2022/Version 7 

LEM reserves the right to carry out modifications on its transducers, 

 

in order to improve them.

LEM International SA 

Route du Nant-d’Avril, 152

1217 Meyrin 

 www.lem.com

ITZ 5000-S FLEX ULTRASTAB

The table shown below summarizes various frequency values and the typical value of over current trip level to detect over current 

state:

Primary current frequency (Hz)

Typical over current trip level to detect over current state

DC (0 Hz))

120 % of 

I

P N

 

DC

1

130 % of 

I

P N

 

DC

1.5

140 % of 

I

P N

 

DC

50

160 % of 

I

P N

 

DC

800

170 % of 

I

P N

 

DC

1600

186 % of 

I

P N

 

DC

Under these conditions:

-  The signal Status (Contact between Pin 4:C and Pin 9:E) switches off, this signal becomes open (No current from collector to 

emitter) to indicate that the unit is not OK.

- The status green LED located on the front panel switches off (fault state)

- The transducer is protected against damage that the over current primary AC may cause to the electronics.

The transducer will automatically starts again to work when the primary AC peak becomes lower than the trip level indicated in 

the table above. 

Transducer head description

Connect the transducer head to the electronics rack using the supplied transducer head cable.

The side of cable where there is the ferrite must be connected to the rack electronics.

The transducer head cable is available in three lengths: 10, 20 and 30 meters.

For more information, see section “ITZ mandatory accessories” on 

page 14

.

Analog output connector

Normally the provided analog output cable should be used. It is also possible to configure your own output cable, bearing the 

following description in mind.

The female 15-pin D-Sub connector Analog out connector contains the following signals:

1. 

Current return

2. 

Current return 

3. 

Current return

4. 

No connection

5. 

No connection

6. Ground

7. 

No connection

8. 

No connection

9. 

Current out

10. 

Current out

11. 

Current out

12. 

No connection

13. 

No connection

14. 

No connection

15. 

No connection

The pins 1 to 3 are internally shorted by rack electronics. It is the same for pins 9 to 11.

Details of further considerations relating to connect the analog output connector directly to a current measuring device can be 

found in the ITZ FLEX ULTRASTAB user guide.

15 

9

  1

1

8

9

15

Summary of Contents for ITZ 5000-S FLEX ULTRASTAB

Page 1: ...9 pin D sub male output connector Secondary current monitoring on BNC connector 4 mm banana sockets for secondary current output Full featured indicator panel Advantages Very high accuracy Excellent...

Page 2: ...required cleaning by dry compressed air is authorized This transducer is a build in device whose hazardous live parts must be inaccessible after installation This transducer must be mounted in a suit...

Page 3: ...ons over voltage cat III pollution degree 2 Rated insulation RMS voltage reinforced insulation UNm V 2000 IEC 61010 1 conditions over voltage cat III pollution degree 2 RMS voltage for AC insulation t...

Page 4: ...ion IP 0 A Mains supply 230 V RMS PC W VA 38 82 Average Peak Power consumption IP IP N DC 5000 A Mains supply 230 V RMS PC W VA 76 116 Average Peak RMS noise current 0 10 Hz 2 Ino ppm 0 02 RMS noise c...

Page 5: ...40 45 50 RM 0 5 RM 1 RM 1 5 Cable length m provided by LEM Primary nominal RMS current versus Cable length TA 0 to 55 C Primary nominal RMS current A Primary nominal RMS current versus cable length Pr...

Page 6: ...000 4000 3000 2000 1000 0 1000 2000 3000 4000 5000 6000 7000 8000 0 5 10 15 20 25 30 35 40 45 50 RM 0 5 RM 1 RM 1 5 Cable length m provided by LEM Primary current measuring range versus Cable length T...

Page 7: ...them LEM International SA Route du Nant d Avril 152 1217 Meyrin www lem com ITZ 5000 S FLEX ULTRASTAB Typical power consumption 0 10 20 30 40 50 60 70 80 90 6000 5000 4000 3000 2000 1000 0 1000 2000...

Page 8: ...d LED This LED is lit when the transducer head saturates due to excessive primary current 600 A 24000 A yellow LED These LEDs are lit when a transducer head with the corresponding full scale range is...

Page 9: ...tatus Interlock signal outputs 9 pin D Sub male UNC 4 40 screw lock 8 Option A B These positions are for future expansion or customization not used 2 8 3 1 7 6 5 4 LEM International SA Route du Nant d...

Page 10: ...is wired it can be Active Low or Active High as shown below In the Diagram A Active Low Output the output signal Uout switches to GND when the corresponding LED is ON In the Diagram B Active High Outp...

Page 11: ...ducer head is not saturated High IP UC The primary current is higher than 105 of nominal full scale output 0 2 V The primary current is lower than 105 of nominal full scale output ACTIVE UC The primar...

Page 12: ...off When the rack electronics temperature cools to approximately 60 C the measuring circuit is automatically re enabled and the status LED is again lit Over current protection Electrical specification...

Page 13: ...AC peak becomes lower than the trip level indicated in the table above Transducer head description Connect the transducer head to the electronics rack using the supplied transducer head cable The sid...

Page 14: ...ransducer head European Schuko CEE 7 7 and US Japanese mains cables 2 m each with three pole IEC C13 female connector Short plug for 4 mm banana terminals and 15 pin D Sub short plug for analog output...

Page 15: ...modifications on its transducers in order to improve them LEM International SA Route du Nant d Avril 152 1217 Meyrin www lem com ITZ 5000 S FLEX ULTRASTAB Dimensions Rack electronics in mm general tol...

Reviews: