FLIM Data Acquisition
Whenever you toggle between these options, the split-off filter on the stand and the X1 port
are automatically moved to the correct position and the corresponding detectors are
activated.
If there are pulsed diode lasers in your system, this selection also impacts the automatic
adjustment of the fluorifier disc (if present), according to
.
The value displayed in the
Count Rate Monitor
is the count rate of the automatically selected
detectors (thus the APDs for external FLIM, the SP FLIM detectors for internal FLIM, and the
HyD RLD for NDD FLIM). Detectors that have been activated but not used display a dark
count rate, whereas detectors that have not been switched on or activated display the
number
0
.
You can change the automatically applied settings later.
FLIM data is collected on the separate SMD workstation using the SymPhoTime software
and the FLIM image is displayed there. In addition, if APDs or SP FLIM detectors are used,
an intensity image is generated in the LAS AF software during the collection of FLIM data.
16.2.1.1
FLIM Data Acquisition with Internal SP FLIM Detectors
For a 5-channel system with a single internal SP FLIM detector, the detector should be at
position 4; two internal FLIM detectors should occupy positions 2 and 4 (
). These
settings have been preconfigured and cannot be changed by the user. SP FLIM detectors
can be used for normal intensity image acquisition (
) as well as FLIM data
acquisition.
Figure 120:
Beam Path Settings in LAS AF for internal FLIM data acquisition. Other detectors may also be
activated at the same time.
Proceed as follows to prepare the TCS SP8 for the collection of SP FLIM data:
1. Open the
Setup
tab.
2. In the
FLIM Detector
pull-down menu, select the
Internal (SP FLIM)
option (
).
The SP FLIM detector should be automatically activated in the
FLIM
mode.
If the SP FLIM detector has been deactivated by the user in the meantime, it can be
activated by clicking the
ON
button (
).
Then the FLIM detector goes into
FLIM
mode. A
Gain
of 990 V suitable for FLIM data
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......