127
FLIM Data Acquisition
16
FLIM Data Acquisition
You can acquire a FLIM image using a pulsed laser, an internal or external FLIM detector,
and an external PicoQuant TCSPC unit.
16.1
Setup Imaging Step – Image Acquisition
1. Start the FLIM Wizard in LAS AF. To do so, open the operating mode selector and select
FLIM
(
).
Figure 110:
Selecting FLIM Wizard
2. Open the
Setup Imaging
step.
The scan mode xt and sequential scan are not supported. Except for this, the image
acquisition procedure is the same as the general procedure outside the FLIM Wizard.
16.1.1
Selecting Detectors for the Image Acquisition
You can use PMTs or APDs for image acquisition. PMTs are distinguished by having wide
dynamic range; APDs are distinguished by higher quantum efficiency.
16.1.1.1
Internal Photomultipliers (Including SP FLIM PMT)
1. In
Beam Path Settings
in the working area, check whether the
Mirror
option is set to
X1
Port
. If not, open the pull-down menu for X1 Port and select
Mirror
(
).
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......