143
FLIM Data Acquisition
Recommended settings of the fluorifier disc for SP FLIM detection:
Table 19: Recommended settings for Fluorifier disc for SP FLIM
As long as the Fluorifier Disc
Auto Select
is active in the dialog window, the position of the
Fluorifier disc is automatically set. It can be changed by the user if
Auto Select
is disabled.
To do so, proceed as follows:
1. In
Beam Path Settings
, click on the
Fluorifier Disc
button (
Figure 125:
Fluorifier Disc button in Beam Path Settings
The following dialog box opens:
Figure 126:
Fluorifier Disc Settings
Excitation laser line
Recommended setting of the fluorifier disc
1
405
Barrier filter 405/640
2
405+470
Barrier filter 405/470
3
405+ 640
Barrier filter 405/640
4
470
Barrier filter 405/470
5
470+640
Barrier filter 470/640
6
640
Barrier filter 405/640 or 470/640
7
MP
SP680 or SP700
8
WLL
Depending on the selected white light laser wavelength,
there are barrier filters for:
488, 514, 561, 594, or 633
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......