5.5
Adjusting the Circular Level of the Prism Pole
1.
Suspend a plumb line.
1
2
4b
4a
TS_080
2.
Use a pole bipod, to align the
prism pole parallel to the plumb
line.
3.
Check the position of the circular
level on the prism pole.
4.
a
If the circular level is
centred, no adjustment is
necessary.
b
If the circular level is not
centred, use an allen key
to centre it with the
adjustment screws.
☞
After the adjustments, all adjusting screws must have the same
tightening tension and no adjusting screw should be loose.
5.6
Inspecting the Laser Plummet of the Instrument
The laser plummet is located in the vertical axis of the instrument. Under nor-
mal conditions of use, the laser plummet does not need adjusting. If an
adjustment is necessary due to external influences, return the instrument to
any Leica Geosystems authorised service workshop.
360°
d
3 mm / 1.5 m
Ø 2.5 mm / 1.5 m
18448_001
1
2
3 - 4
The following table explains the most common settings.
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Turn on the instrument. Using the tribrach footscrews, level the
instrument with the electronic level.
☞
Electronic level:
•
Select
Devices
from the Home Menu.
•
Tap the arrow button to the right of the device name.
•
To display the electronic level, tap
Compensator
.
Adjusting the circular
level step-by-step
☞
Inspecting the laser
plummet step-by-step
Check & Adjust
41