61
Fig. 60
107/2 lamp housing, opened
1
Collector,
2
Lamp holder with 12 V/100 W halogen lamp
Fig. 59
107/2 lamp housing
1
Screw for opening of lamp housing
Replacement fuses in compliance with IEC 127-2
and/or UL 198 G and/or manu-facturer type:
Part no:
846-205.000-00
Designation:
T 4A
Wickmann 19 195/
Schutter FST
Note
Never use replacement fuses with a different
current rating.
• Connect the microscope and the power unit to
the mains supply.
Replacing the 12 V/100 W halogen lamp in the
106, 107, 107/2 lamp housings
Ask a Leica field technician to show you the pro-
per procedure for replacing the halogen lamp.
All necessary steps are listed below.
Attention!
Prior to making any assembly work, always
disconnect the power supply at the external
transformer outlet and the microscope out-
let!
• Switch off the microscope and the power unit.
• Disconnect the power cords of the micro-
scope and of the power unit.
• Unscrew the lock screw on the microscope
and remove the lamp housing.
Fig. 61
106 lamp housing, opened
1
Screw for opening of lamp housing,
2
Lamp holder with
12 V/100 W halogen lamp,
3
Collector,
4
Diffuser disc
4
3
2
1
1
2
1