Leica DM750 P User Manual
Get Ready! 24
Substage Condenser (Continued)
4. Completely loosen the two thumbscrews
on the condenser mount.
5. Slide the substage condenser into the
fork mount located underneath the stage
by aligning the positioning pin on the
condenser into the slot on the back of the
fork.
6. Raise the condenser to its highest position
using the condenser focus on the left-hand
side of the substage
7. Tighten the thumbscrews to roughly center
the substage condenser so that the top lens
of the condenser is centered beneath the
objective in use. You will more accurately
center and focus the condenser when you
reach section Koehler Setup.