background image

Extended Measurement Information 
 
The SMART key can be used to show extended information about a measurement. Repeated 
key presses cycle through the following parts. After the last part, the display returns to the 
standard contents. 
 
Normal Measurements: 
1) Azimuth, horizontal distance, vertical distance, total distance. 
2) Roll angle and dip angle (inclination of the magnetic field). 
3) Magnitude of the magnetic field and the acceleration/gravity. 
4) Raw sensor values (x,y,z) of the on board acceleration sensor. *) 
5) Raw sensor values (x,y,z) of the separate acceleration sensor. *) 
6) Raw sensor values (x,y,z) of the magnetic field sensor. *) 
 
Calibration Measurements: 
1) Combined acceleration sensor values (x,y,z). 
2) Magnetic field sensor values (x,y,z). 
3) On board acceleration sensor values (x,y,z). *) 
4) Separate acceleration sensor values (x,y,z). *) 
 
*) These parts are not available when reading from memory. 
 
 
Bluetooth Connection 
 
When Bluetooth is switched on, the device is discoverable and connectable at any time. It 
appears as “DistoX-nnnn” where nnnn is the serial number of the device. It provides a serial 
connection (SPP) named “serial”. The device needs no pairing key. If the connecting device 
request one, it is “0000” (four zeroes). 
 
To read the results from the device, a special program is needed (PocketTopo). This runs 
either on a PC or on a PDA. To Allow the program to connect to the Disto, the Bluetooth port 
must be selected under Menu:Options:Port. The port is listed in the Bluetooth manager as the 
“outgoing serial port” for the device. 
 
When Bluetooth is switched on, a Bluetooth symbol appears at the top of the display. The 
symbol flashes whenever another device is connected. Except when a memory entry is shown, 
the display shows the number of unsent measurements in the top right corner. 
 
It is possible to switch the Laser on or off over Bluetooth and to remotely trigger a 
measurement. The corresponding commands are in the Bluetooth menu. 
 
The Disto does not automatically switch off as long as there is an active Bluetooth 
connection. 
 
 
 
 
 
 

Summary of Contents for DistoX2

Page 1: ...milarly to the original X310 The display shows the Azimuth in the first the inclination in the second and the distance in the bottom line It stores up to 1000 measurements to be inspected later or transferred over the Bluetooth connection Main Functions DIST power on start Laser measure distance CLR cancel current operation switch Laser off REF change distance reference TIMER start timer automatic...

Page 2: ... 180 is added to the azimuth On the display a down arrow is shown near the Laser symbol when the option is turned on and whenever a backsight measured memory value is shown Triple Shot Check If the triple shot check option is turned on the device checks for three nearly identical measurements in a row If a triple is detected the device beeps twice and shows an identical sign to the right of the me...

Page 3: ...luetooth is switched on the device is discoverable and connectable at any time It appears as DistoX nnnn where nnnn is the serial number of the device It provides a serial connection SPP named serial The device needs no pairing key If the connecting device request one it is 0000 four zeroes To read the results from the device a special program is needed PocketTopo This runs either on a PC or on a ...

Page 4: ...creen to the next Use the SMART key to return to the previous screen Screen 1 Battery Shows the actual battery voltage and the battery chemistry used for the battery level indicator LI for LiPo or AL for Alkaline Press the FUNC and SMART keys for 5 seconds to change the battery chemistry Be sure to use the correct setting LI default for a built in rechargeable battery or AL for AAA primary cells S...

Page 5: ...ark the rear exit point of the Laser beam picture Place this point as near as possible to the survey point Switching the display illumination off completely saves a lot of power For best performance the device should be calibrated in regular intervals Error Codes If an error occurs during a measurement the text Info appears on the screen together with one of the following error codes 252 Temperatu...

Page 6: ...proper calibration Features Selectable Units m ft inch grad Memory capacity 1000 measurements Laser Type 635nm 1mW class II Mechanical Size 55 x 31 x 122mm Weight 150g Protection IP65 Electrical Battery voltage 1 5 5 5V LiPo charger input 4 5 6V Power consumption Bat 4V Bat 3V original X310 Bat 3V Idle incl Bluetooth 4mA 5mA 36mA Illumination on 9mA 11mA 43mA Laser on 88mA 110mA 120mA ...