23
Leica CM1860/CM1860 UV
Installing the Instrument
4
4.2 Unpacking instructions for the Leica CM1860/CM1860 UV cryostat
Note
•
When the instrument is delivered, check the tilt indicators
on the packaging.
• If the arrowhead is blue, the shipment was transported laying flat, was tilted at too great an angle or
fell over during transport. Note this on the shipping documents and check the shipment for possible
damage.
•
Two persons are required to unpack the instrument.
•
The illustrations serve only as an example to explain the unpacking procedure.
Fig. 3
Warning
Caution when removing the strips
! There is a risk of injury when opening them (the strip
has sharp edges and is under tension)!
•
To remove the strips
, suitable shears and safety gloves are required.
•
Stand next to the packaging and cut through the strips at the locations shown
•
Lift up the covering box
and remove it.
•
Carefully remove the adhesive tape
holding the two transport anchors
both sides of the instrument and remove them.
•
Then remove the dust cover
from the instrument.
•
Remove the two white and the two blue transport anchors
, which protect the cryochamber
window.
•
•
Lift and remove the wooden edge
•
Remove the ramp
on the front side of the pallet
and properly insert it into the
rear side
.
•
Insert the ramp properly. Make sure the ramp components designated "
L
" (left) and "
R
" (right) click into
place in the provided guide channel
•
When correctly installed, the guide rails
are located inside and the arrows
point toward each other.