77
Leica CM3600 XP
Cleaning and Maintenance
8
•
Remove the knife from the knife holder, see
.
•
Lock the window, see
•
Open the chamber lid, see
.
1
2
3
4
5
6
6
7
8
1
6
2
7
3
8
4
5
6
Fig. 87
•
Remove the screws
. When removing the screws
, make sure that the spacer rings
between the bracket
the microtome
do not get lost.
•
Fold down the belt covers
.
•
Disconnect the connections
on the left side of the cabinet (not on the microtome).
• Fasten cables or straps that have a sufficient carrying capacity to the carrying handles
.
•
Carefully lift the microtome
out of the cabinet using the crane.
•
In doing so, securely fasten the microtome
to prevent rotation or vibrations as this could
lead to damage to the cabinet.