39
Leica CM3600 XP
Software
6
Set chamber temperature
Click on this display to change the currently selected value. Temperature values from -30 to +1 °C are
permitted.
Alarm temperature
Click on this display to change the currently selected value. Temperature values from -25 to -5 °C are
permitted.
Note
The alarm temperature should not be too close to the set chamber temperature. If the alarm
temperature is too close to the set chamber temperature, an alarm can be triggered simply when
working with the instrument during a prolonged period of time or during the daily defrost cycle.
A temperature difference of approx. 5 K (Kelvin) is recommended.
Time
and Date
Fig. 29
Displays time and date. Clicking on this display opens the
System time
dialog box
.
•
Clicking on a subarea of the display (day, month, year, hour or minutes)
lets you change the setting.
• The entry must be confirmed by pressing the
Ok
button.
Start time 1 of the defrost
Fig. 30
Displays start time of the 1st automatic defrost cycle. Clicking on this
display opens the
Defrost time 1
dialog box
•
Click on individual parameters (hours or minutes) to change the
corresponding settings.
• The entry must be confirmed by pressing the
Ok
button.
Start time 2 of the defrost
Fig. 31
Displays start time of the 2nd automatic defrost cycle. Clicking on this
display opens the
Defrost time 2
dialog box
•
Click on individual parameters (hours or minutes) to change the
corresponding settings.
• The entry must be confirmed by pressing the
Ok
button.
Clicking on this button activates or deactivates the 2nd automatic defrost
cycle.