11
VT1000 S
Instrument Characteristics
3
3.
Instrument Characteristics
3.1
Technical data
General data
Sectioning frequency (± 10 %)
0 - 100 Hz
Amplitude
adjustable in 5 steps: 0.2; 0.4; 0.6; 0.8; 1 mm
Sectioning speed (± 10 %)
0.025 - 2.5 mm/s
Return stroke speed (± 10 %)
5 mm/s
Total vertical specimen stroke
15 mm (motorized)
Sectioning range
1 - 40 mm (adjustable)
Specimen retraction
0 - 999 μm (adjustable; can be deactivated)
Maximum specimen size:
with standard knife holder
33 x 40 mm
Specimen orientation
330°
Section thickness selection
1 - 999 μm, in 1-μm steps
Magnifier, assy. (standard accessory of the
configured instrument)
2 x magnification
Ambient conditions
Operating temperature range
min. 5 °C to max. 40 °C
Relative humidity
max. 80 %
Height
up to 2000 m above sea level
Electrical data
Rated voltage range (± 10 %)
100 V - 240 V
Nominal frequency (± 10 %)
50 - 60 Hz
Power draw
50 VA
Mains fuse
T 1.25 A
Pollution degree
2
Overvoltage installation category
II
Overload protection
yes
Internal current limiter of electronics
yes
Dimensions
L x W x H
480 mm x 360 mm x 200 mm
Height with magnifier support
285 mm
Weight:
(without magnifier support)
17 kg
(magnifier support only)
2 kg
(total)
19 kg