WaveAce Series
47
WA-GSM-E RevB
1.
Press the
Type
option button and select
Pulse
.
2.
Press the
Source
option button to choose a channel input (CHS 1 - 2;
or CHS 1 - 4 on 4 channel models) or external input.
3.
Press the
When
option button to select how to compare the trigger
pulse relative to the value selected in the Set Width option.
4.
Press the
Set Width
option button and then turn the
Adjust
front
panel knob to set the pulse width.
5.
Press the
Next Page
option button.
6.
Press the
Mode
option button to select
Auto
,
Normal
, or
Single
mode. Use
Auto
mode to let the acquisition automatically run in the
absence of a trigger. Use
Normal
mode when you want to see only
valid triggered waveforms (when you use this mode the oscilloscope
does not display a waveform until after the first trigger). Use
Single
mode when you want the oscilloscope to acquire a single waveform.
7.
Press the
Setup
option button to display the
Trigger Setup
menu.
You can use the
Trigger Setup
menu to select a coupling mode and
define a
Holdoff
value.
Coupling
modes comprise
DC
,
AC
,
HF Reject
,
and
LF Reject
. Turn the
Adjust
front panel knob to set the
Holdoff
value.