Calibration
Focus II
System Guide
53
Attenuator Calibration - applicable to calibration of Strain Gage and
Thermocouple modules
The first step is to measure the DC offset of the attenuator Output
channel. A dialog box will appear. The operator should read the DC
measurement from the voltmeter and type the value in the edit box.
Please allow enough time for voltmeter readings to stabilize. Stability
can be assumed if the readings vary less than 2% of the total
measurement reading on all ranges.
DC Measurement for Attenuator Offset Calibration
You will also be requested to measure the RMS value of the Attenuator
signal source and type it into the dialog box.
Note: The voltmeter
should remain in DC measurement mode.
DC Measurement for Attenuator Gain Calibration
Summary of Contents for Dactron FOCUS II