Calibration
Focus II
System Guide
52
Signal Output Calibration Verification
After the signal output is calibrated, the software verifies the calibration
results. During this verification, it requires measuring again the DC
offset and RMS values.
A dialog box will ask for a measurement and entry of the DC reading
from the voltmeter. The software will check the deviation of its expected
value and the measured value.
DC Measurement for Offset Verification
DC Measurement for Gain Verification
After verification, both the offset and gain error should be within
tolerance.
The pass/no-pass tolerance used during the verification is:
10 volt range: 1% of the Full Scale
Once this step is completed, the operator will be prompted to repeat this
process for Output 2.