Table 11.1 Rotors and adapters
Rotor
type
ID
code Tube/bottle Adapter
Maximun
speed
(rpm)
Maximum
RCF
(×g)
Allowable
imbalance(*)
imbalance
volume
imbalance
MCEN-
A05-
001
01
1.5/2.0ml
tube
15000
15100
2.0g/
tube
5mm/
tube
0.2ml
PCR tube
MCEN-
A06-
024
15000
15100
0.5ml
microtube
MCEN-
A07-
024
15000
15100
11.1.3
Notice
◼
The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated
samples density is over 2.0g/ml, please calculate allowable speed depending on the
following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
11.1.4
Autoclaving
◼
The rotor makes by high-strength aluminum alloy material, can be high-pressure sterilization:
121
℃
(
1.0kg/cm
2
)
20 minutes.
CAUTION: The lid of the rotor is made of plastics, can not be high-pressure
sterilization, only ordinary sterilization can be used.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes, please refer to the table below.
Table 11.2 Cleaning and sterilizing conditions for tubes
O
:
Applicable X
:
Inapplicable
Condition Material
PA
PC
PP
C
leanin
g
Cleaning
fluids
Acidic
(
pH5 or lower
)
X
X
X
Acidic
(
higher than pH5
)
O
O
O
Alkaline
(
higher than pH9
)
O
X
O
Alkaline
(
pH9 or lower
)
O
O
O
Neutral
(
pH7
)
O
O
O
Warm water(up to 70
℃
)
O
O
O
Ultrasonic
cleaning
Neutral detergent
(
pH7
)
O
O
O