Figure 6-2 the main interface
The main interface as figure 6-2. At this point speed was set to be 15000rpm, the running time
was 12 minutes. When speed symbol
is rotating, indicating the centrifuge is running. Time display
symbol
display the rate of working to time setting. The total time setting is divided into 10 scales.
7. Rotor Preparation
7.1 Prepare the sample
7.2 Inject the sample into tubes.
CAUTION
:
Excessive sample may cause leakage in the centrifuge tube, keep the
actual capacity.
◼
Do not exceed the actual capacity specified in the instruction manual.
7.3 Keep the tubes balance
◼
See Table 7.1 for allowable imbalance of each rotor.
◼
Although this centrifuge can accept sample balancing by eye, we recommend that you use this
centrifuge in a well-balanced condition to prolong its life expectancy.
◼
Never intentionally run the centrifuge under unbalanced condition even though the allowable
imbalance is not exceeded.
7.4 Inspect the rotor
CAUTION
:
●
Any abnormality such as corrosion or scratches are found, stop using the rotor and
contact our service center.
●
Only manufacturer
’
s rotors which claimed can be used with the unit are allowed.